👉 CIRCUIT CONSTRUCTION
Activities
- Construct circuit on the breadboard to test the sound sensor and to get preliminary result.
- The equipment used for this test is Terumo Terufusion TE-311 syringe pump.
- The sound sensor captured the alarm sound from the syringe pump in the condition where the syringe pump having plunger error and condition where low battery.
- The alarm sound showed the wave pattern in the oscilloscope where I able to know the frequency and amplitude of the alarm sound.
Example of condition syringe pump battery low:
Frequency: 4.108kHz
Amplitude: 2.52V
Example of condition syringe pump plunger error:
Frequency: 3.654kHz
Amplitude: 2.48V
Frequency: 3.737kHz
Amplitude: 2.4V
Frequency: 3.903kHz
Amplitude: 2.48V