WEEK 15
DATE: 04/11/2019-08/11/2019
SUPERVISOR: SIR SHAH RIZAILLI BIN MUKHTAR

👉 MEETING WITH SUPERVISOR
  • The meeting was held in FYP room level 2. The purpose of this meeting was to update the improvement that have been done of the prototype to the supervisor. The supervisor verified the coding, programming and also the functionality of the device on the real medical equipment.
  • The result of the meeting were:
  • 1) The student need to improve the knowledge of the programming (Arduino software).
  • 2) The student need to do repairing of the circuit so that there will be no distraction during collecting the data.
  • 3) The student need to collect more data regarding alarm failure pattern to be included in the technical report.
👉 DATA COLLECTION 
  • The student continue to collect data in FYP room. The data collection includes the battery failure of syringe pump and also battery failure of patient monitor. The battery failure of syringe pump have 5 stage but I only collect 4 alarm pattern. But, for patient monitor, it only have 1 battery failure alarm.  
  • Below is one of the example on how I collect the data for the stage 1,2,3 and 4 of syringe pump battery failure. 


  • For stage 1, the frequency measured for the ON pulse is 1.190Hz. For stage 2, the frequency measured is in the range of 2. 381Hz while stage 3 is 4.762Hz. Lastly, frequency measured for stage 4 ON pulse is 9.259Hz. This result shows that the frequency for each stage is different and it also have different length of ON and OFF time. 
  • For the next week, I will be collecting more data and get the average frequency for each stage.